- This system ensures consistent dispensing by integrating a ring pump powered by a DC motor with an encoder and a dedicated controller (RE-C300), which precisely halts the motor at integer multiples of a full rotation.
- The maximum dispensing volume is 150 mL, with increments of 5 mL per step.
- Available tubing options include three types: Silicone, Tygon A-60-F, and Transmaster TM-15.
Note: Details such as specifications, etc., may be changed without notice.
MOTOR CONTROLLER - RE-C300
- A dedicated motor controller designed specifically for the constant discharge pump system.
- Install the included software from the USB memory onto your computer. By connecting the computer to the RE-C300 via the included USB cable, you can easily control various pump operations.
- The system allows for programming of multiple pump operation sequences.
- The motor controller can operate independently, offering stand-alone control and functionality.
| Model Number | RE-C300 | 
| Outer Dimensions | 180 × 150 × 42 mm | 
| Drive Voltage | 24 VDC (AC / DC Adapter Included) | 
| Applicable Pumps | RP-MC Series RP-HC Series
 RP-HⅡC Series
 RP-KⅡC Series
 | 
| Control Method | Original system for our fixed discharge volume pump system. | 
| Accessories | AC/DC Adapter USB Connecting Cable (Type-A to Micro-B)
 Pump Connection Cable
 USB Flash Drive (Containing Software)
 | 
 
Note: Details such as specifications, etc., may be changed without notice.
MECHANISM OF THE RING PUMP




Note 1: Details such as specifications, etc., may be changed without notice.
Note 2: Visual diagram above is from Aquatech Co., Ltd.
		
		
		
		
			SPECIFICATIONS
			Swipe left to see the remaining contents
		
		
			
				
				
				
				
				
				
					
						| Type | Ring Pump | 
				
				
              
				
				
				
				
				
				
				
				
				
					
						| Rated Voltage | 24 VDC | 
				
				
				
              
				
              
				
				
				
				
					
						| Tube Material | Silicone, Tygon®A-60-F, or Transmaster TM-15 | 
				
				
				
				
				
				
				
				
				
				
				
				        
        
				
				
				
				
				
              
				
              
				
              
				
              
				
              	
              
				
              
				
              	
              
				
              
				
              	
              
				
              
				
				
					
						| Notes | Maximum Discharge Volume - 150 mL*1 
	
		
			| Constant Discharge Pump System
 Model #*2
 | Voltage | Tubing Material | I.D. | O.D. | System Components |  
			| - Pump Model # - Controller Model #
 |  
			| RE-C300-KⅡC05S | 24 VDC | Silicone | 4.8 mm | 8.0 mm | - RP-KⅡC05S-150Z-DC24VE - RE-C300
 |  
			| RE-C300-KⅡC05N | Tygon®A-60-F | 4.76 mm | 7.94 mm | - RP-KⅡC05N-150Z-DC24VE - RE-C300
 |  
			| RE-C300-KⅡC05W | Transmaster TM-15
 | 4.8 mm | 7.9 mm | - RP-KⅡC05W-150Z-DC24VE - RE-C300
 |  Note: Details such as specifications, etc., may be changed without notice. *1: The maximum discharge volume refers to the volume per single operation that can be set on the RE-C300. (The maximum number of rotations that can be configured for one operation is 99 rotations.) *2: The first unit will be sold as a complete "Constant Discharge Pump System Model Number," including both the pump and the motor controller RE-C300. Subsequent purchases will allow the pump to be sold as a stand-alone unit. *:"Tygon®" is a trademark of Saint-Gobain Performance Plastics. |